| OK (Pass) evaluation |
|
This implies that the problem is not with
the tester. Please follow the steps listed below for further
instructions.
|

| Are the Jigs in satisfactory condition? |
|
Please confirm the status of the jigs for
any fractures or other deformations. Replace if
necessary.
|

| What is the status of the work? |
|
Replace the work with a reliable master or
another work piece and conduct testing.
|
|
|
 |
|
OK (Pass) evaluation. 1. The material, form or
processing is faulty. 2. Possible cause for instability: Temp. variation or
temp change
|
|
| |
|
|

|
|
| Extend charge time. |
|
Extend charge time to 3X the current
setting.。
|
|
|
 |
|
Decreased detection value Possible cause: Temperature
influence and deformation of the jigs or work parts within the test
circuit
|
|
| |
|
 |
|
Unchanged detection value Implies a leak in the jig,
work or piping structure.
|
|
|
|
| NG (Fail) evaluation. |
|
The problem is within the tester. Asses the
probable cause and contact your Fukuda agent for further maintenance
details.
|

| Extend charge time. |
|
Extend charge time to 3X the current
setting.
|
|
|
 |
| Decreased detection value |
|
The tester's pneumatic parts are
deteriorated. Pneumatic parts overhaul or replacement parts are
necessary.
|
|
| |
|
|
|
| Unchanged detection value |
|
Set the test pressure to 0 (zero) and
conduct testing.
|
|
|
 |
|
Use the regulator and conduct testing by setting the test
pressure to 0 (zero).
|
|
| |
|
 |
|
OK (Pass) evaluation. The tester's pneumatics are
faulty 1. Sensor Overhaul 2. Clean or replace solenoid valve 3. Clean
or replace internal
piping
|
|

NG (Fail) evaluation The tester's
electrical system is faulty. 2. l Readjust or replace the circuit
board |
|