AIR LEAK TESTER CALIBRATION & INSPECTION & PREVENT MAINTENANCE

 


Fukuda Test
Environment Solution

Fully Calibration

  • Routine and complete annual 13 point inspections for Air Leak Tester equipment.
  • A written inspection report after each inspection indicating what has been checked.
  • Notification of any apparent wear of parts and visible corrosion and recommendations for efficiency and reliable cooperation. With this report you will better be able to plan for any necessary repairs and insure proper upkeep through routine maintenance.
  • A 10 % discount off of list price for part replacement within the Air Leak Tester equipment .
  • Reliable tech support at no service call charge.

13 Point Inspection

  • Check the external appearance for corrosion and damage.
  • Check the air filter for dirt
  • Check the air lines piping for leak or damage
  • Check the switches and indicator lamp
  • Test for sequential operation (OK , small leak and large leak judgment) Validation of detection sensitivity judgment
  • Check for timer synchronization
  • Check for value operation
  • Check for zero point correction
  • Check and ensure appropriate air pressure
  • Calibration of differential pressure and gauge pressure
  • Testing unit for stability Confirmation of good sample (master)